USER INTERFACE
COMBINATION OPERATION
The Combo modality allows the user to select and use ultrasound therapy in combination with electrical muscle stimulation.
Combination therapy utilizes the Ultrasound modality in conjunction with High Voltage Pulsed Current (HVPC), IFC
Premodulated (2p), Asymmetrical Biphasic, Symmetrical Biphasic, or VMS
In this mode of therapy, the Ultrasound Applicator becomes one half of the electrical circuit. An electrode attached to the Red
(+) Lead Wire completes the circuit .
Complete the following steps to begin Combo treatment:
1. Prepare Patient and therapy system
patient, and securing electrodes. Ultrasound Patient preparation is found on page 32.
2. Connect the Red (+) Lead Wire from Channel 1 to the electrode.
electrode . The Black (-) Lead Wire is not used. The Ultrasound Applicator completes the circuit for Combination Therapy.
3. From the HOME SCREEN, select the COMBO icon.
4. Select the ultrasound combination therapy desired by touching the corresponding icon.
Note: for safety reasons not all wave forms are available for combo therapy.
5. SET UP TREATMENT
On the treatment review screen you can adjust treatment parameters to desired level.
Never start with intensity adjustment – first adjust all other parameters and set Intensity just before starting treatment
Pressing the save button will save the treatment as a custom protocol which can be named by the user
Pressing the Assign to button will give two buttons as seen below
- Assign to: this button assigns the treatment data to a specific treatment data file which can be named by the user
- Open pain scale: this button opens up the pain scale so the pre-treatment pain can be recorded
INTELECT
LEGEND 2 USER MANUAL
®
– Refer to the PATIENT PREPARATION section for electrode selection, preparing the
to generate a therapeutic effect .
Make certain the Lead Wire is completely seated in the
EN