ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES (CONTINUED)
The Intelect® Legend 2 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Intelect® Legend 2 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic
± 8 kV contact
discharge (ESD)
± 15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply
transient/burst
lines
IEC 61000-4-4
± 1 kV line(s) to line(s)
Surge
IEC 61000-4-5
± 2 kV line(s) to ground
Voltage dips, short
0% UT; 0.5 cycle at 0°,
interruptions and
45°, 90°, 135°, 180°, 225°,
voltage variations
270°, 315°
on power supply
0% UT; 1 cycle and 70%
input lines
UT; 25/30 cycles Single
IEC 61000-4-11
phase at 0°
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
Proximity magnetic
CW; 8 A/M
fields
2.1 KHZ; 65 A/M
IEC 61000-4-39
50 KHZ; 7.5 A/M
NOTE UT is the a.c. mains voltage prior to application of the test level.
INTELECT
LEGEND 2 USER MANUAL
®
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply
lines
± 1 kV differential mode
± 2 kV line(s) to ground
0% UT; 0.5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270°, 315°
0% UT; 1 cycle and 70%
UT; 25/30 cycles Single
phase at 0°
30 A/m
CW; 8 A/M
2.1 KHZ; 65 A/M
50 KHZ; 7.5 A/M
Electromagnetic Environment - Guidance
• Maintain humidity in the use environment to at
least 50% relative humidity.
• Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, DJO
recommends implementing additional controls to
maintain relative humidity to at least 50%.
• Communicate these ESD-precautionary
procedures to healthcare staff, contractors,
visitors and patients.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Intelect® Legend 2 requires continued
operation during power mains interruptions, it
is recommended that the Intelect® Legend 2 be
powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Test frequency 30 kHz
Test frequency 134.2 kHz
Test frequency 13.56 MHz