SYSTEM
WAVEFORMS (CONTINUED)
TENS - HAN
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
The HAN Waveform provides optimal parameters with a
precisely controlled sequence of Dense-and-Disperse
(DD) modes of stimulation where a burst of 8 pulses
at 80Hz is alternating with continuous stimulation (no
burst), each lasting for 3 seconds.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Burst Frequency
Frequency
Phase Duration
IRMS
DC component
Microcurrent
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
Microcurrent is a monophasic waveform of very low
intensity.
Output Mode
Available on channels
Treatment Time
Mode Selection
Output Intensity
Duty Cycle
Frequency
Polarity
IRMS
DC component
INTELECT
LEGEND 2 USER MANUAL
®
Electrodes
1, 2, 3, 4
1-60 min
0-100 mA (CC)
2 bps
80 pps
180 μsec
0-19mA
Electrodes
1, 2, 3, 4
1-60 Min
0-1,000 μA
0.1-1,000 pps
Positive, Negative, or Alternating
0- 1mA
VMS
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
VMS is a symmetrical biphasic waveform with a 100 µsec
interphase interval. Because the pulse is relatively short,
the waveform has a low skin load, making it suitable for
applications requiring high intensities, such as in muscle
re-education protocols.
Output Mode
Available on Channels
Treatment Time (Stim)
Treatment time (Combo)
CC
Mode Selection
Output Intensity
Anti-Fatigue
Channel Mode
Cycle Time
Frequency
No
Phase Duration
Ramp
Set Intensity
Setting in Reciprocal and Co-Contract modes
IRMS
DC component
CC
50%
No
Electrodes
1, 2, 3, 4
1-60 min
1-30 min
CC or CV
0- 140 mA (CC) 0-140 V (CV)
Off or On
Single, Reciprocal, Co-Contract
Continuous or User Defined
1-200 pps
30-1,000 μsec
0-5 sec
Individual/both Channel Intensity
0-50mA