18. Speed/RCF setting button
19. Rotor→Time(min./sec.)→Acceleration/Deceleration selection button
20. Second LED
7-1. Minute LED
21. Relative centrifugal force (RCF) LED
8-1. Speed (rpm) LED
22. Deceleration rate LED
9-1. Acceleration rate LED
23. Rotor (adapter) number display
24. Time display
25. Speed/RCF display
26. Acceleration/Deceleration rates display

5.2 Turn on

Connect one end of the cable supplied in the socket of the centrifuge placed at the back
side of the equipment and the other end to the power net (AC230V 50HZ/60HZ) properly
grounded.
After connecting the equipment, press the switch at the back side of the equipment. With a
short buzz, the LED display on control panel will turn on and after a complete auto-test the
equipment will be ready to go to next point.

5.3 Open the lid

Press the lid open button
automatically open; lift the lid manually up to a certain height where it can be held by the spring
and hinges and making accessible the inner chamber.
Note: When pressing the lid open button, make sure not to have the head or other objects
above the lid to avoid hits!
In case of failure and the button does not work, it will be necessary to open the lid
manually according to point "Troubleshooting".

5.4 Close the lid

Push the lid downwards until the hook at the front of the lid passes the lock pin with a
"click" sound; then the bottom of hook will touch the trip switch and the lid will be locked.
Push the lid downwards properly; do not overexert or the
hook will be damaged.
placed on the control panel and the lid will be
- 39 -