VisionAire™
2.6
Safety Features:
situated in the stator winding (135°C / 275 °F). A safety relief valve
Compressor Motor: Thermal safety is ensured by a thermostat
is fitted to the compressor outlet and is calibrated to 280 kPa (40
psig).
Power Failure: In the event the unit is operating and a loss of power
occurs, an audible single – pulse intermittent alarm will activate.
Oxygen Monitor: The oxygen monitor detects any drop in
concentration below 82% (±3%) and activates a visual alarm (amber
LED indicator on front panel). If the concentration continues to
operate below 82%, an audible 2 – pulse intermittent alarm will also
activate.
Hi Pressure: An over-pressure safety feature prevents potential
damage to device components at a pressure of 228 kPa (33 psig (±
2 psig)). An audible 4-pulse intermittent alarm will activate.
Low Pressure: An audible 3- pulse intermittent alarm will activate if
the device pressure reduces to ≤ 35 kPa (5 psig (± 1 psig)).
No Flow Warning: An obstruction to the flow of oxygen, such as a
pinch or kink in the delivery cannula, will activate an audible 6- pulse
intermittent alarm.
Product Filter: ≥ 10 µm filter
MN137-1 rev. H