8.
Assembling
8.1
Installing the objectives
1. Turn the coarse focusing knob ① until the nosepiece reaches
its lowest position.
For a safe transport, the nosepiece is placed in the low-
est position and the tension adjustment collar ② is ad-
justed to the proper tension when the microscope leaves
the factory. (Fig.1)
2. Screw the lowest magnification objective on the nosepiece
from the right side, then turn the nosepiece clockwise. Mount
the other objectives in the same way, following the sequence
from low to high.
Note: the objectives can also be installed through the
stage opening. (Fig. 2)
Clean the objectives regularly. In inverted microscopes, the
objectives are very sensitive to dust.
To prevent dust and contamination from entering the micro-
scope, cover all the unused holes with dust caps ③. (Fig. 3)
When operating, use a low power objective to search and
focus the specimen, then switch to higher magnifications.
When switching between objectives, slowly turn the nose-
piece until it clicks. The click means that the objective is in
the right position, in the center of the light path.
8.2
Installing stage extension or mechanical stage
Stage extension is an optional accessory.
Mechanical stage is an optional accessory for IM-300LD2.
Stage extension can be installed on either side of the
stage to enlarge the working surface.
Mechanical stage can only be installed on the right side.
1. Installing the units: screw the bolts in the fixing holes of the
stage, then mount the unit from below the stage. (Fig. 4)
NOTE: The stage has a series of holes in the underside.
To install the units it is necessary, starting counting from
the front of the microscope, to use the third and fifth
holes. By using a different set of holes, the units will not
be installed correctly.