Annual Inspection
EN
Fig. 8-4
Vacuum pump exhaust ports
32
8.7

Vacuum Pump Operational Test

This test is not designed as a performance test to measure the vacuum pressure
but only to assess if vacuum pumps are operational.
1. If Basic Function Tests 8.3 to 8.5 have been conducted, skip to step 2. If not,
see Basic Function Tests 8.4 steps 1 to 9.
2. Check canisters to be sure at least one port is open.
3. Press the Pause/Resume button.
4. Place a finger adjacent to one or both of the gold exhaust ports (Fig. 8-4, (1))
on the back of the pump and feel for air flow.
Record results in the test log in Section 15.1. Test is successful if air flow is ob-
served.