SETTING UP THE NURSING HOME BED (SAFETY CLEARANCES)

The bed must be set up in such a way that when implementing all adjustment
functions of the bed, collision of bed parts with other objects in the room or
with parts of the room is not possible (e.g. mains cable, windowsills, slanting
ceilings, pipes, surface-mounted installations, wall boards etc.). A safety clea-
rance from walls and other objects of at least 2.5 cm should be maintained in
all possible bed positions when doing this. Caution: Equally no collision may
occur for tilting and height adjustments! The bed must be set up in a way it
can be easily disconnected from the mains supply (by unplugging the mains
plug).

CLEANING AND DISINFECTION OF THE BED

The nursing home bed must be cleaned and disinfected regularly, and at least
before every change of resident. The nursing home bed may not be cleaned
or disinfected in automatic washing units, nor with splash water, jet water or
water spray. Only spray and wiping disinfection is permitted.
The following agents are recommended for the disinfection of the nursing
home bed, and may only be replaced by equivalent agents:
Terralin, Perform and Sagrotan Med (Schülke & Mayr GmbH, Norderstedt).
These agents may only be used in accordance with the manufacturer's regula-
tions! These refer particularly to the dosage of the agents and possible hazards
in combination with other agents.
The wooden linning must not come into contact with these agents.
For particularly stubborn stains, commercially available, environmentally friendly
all-round household cleaners can of course be used.
No cleaners or cleaning agents may be used, however, that contain mechani-
cally abrasive components and that have the effect of scouring agents.
and veneer surfaces when in use and when being cleaned. Wood is a natural
product which has the property of absorbing moisture, and which expands in
doing so. This can cause damage to the surface sealing. Damage to varnished
ask Wissner-Bosserhoff Customer Service.
Manual remote controls may only be cleaned with cleaning cloths for PC key-
boards. If dampness has penetrated, if the unit is heavily soiled or if the keys
stick due to dirt particles, the hand controls must be replaced if necessary.