wissner-bosserhoff Carisma Manual De Instrucciones página 96

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 302
Cleaning the bed may only be carried out in accordance with the specifications
in Section "Cleaning and Disinfection" of these instructions for use, since in-
creased risks could otherwise occur for which the manufacturer can take no
responsibility.
If incontinent patients occupy the bed, an incontinence protection for the
mattress must always be used.
If immobile patients/residents lie in the bed permanently, bed sores can
occur if additional lying aids (e.g. pressure sore mattresses) and other special
nursing measures for pressure sore prevention are not employed. Under no
circumstances is this the responsibility of the manufacturer of the bed.
The product is explicitly not suitable for permanent use. If the specified opera-
ting time is exceeded, the power unit can heat up, causing it to switch off auto-
matically. The power unit must then cool down for at least two hours, and can
only be put into operation again after this time has expired.
It is absolutely essential that a blockage of the bed or of parts of the bed
mechanics is avoided, since this can lead to damage and to the total failure of
the power unit through overheating.
It is absolutely essential that the safe working load is not exceeded, since this
can lead to damage and to the total failure of the power unit through overhea-
ting.
32) Servicing notes – Technical Safety Inspections – Notes on Rectifying
Errors:
Electrically-operated hospital and nursing home beds are medical products
dance with the Law on Medical Products (MPG) and the Medical Products
Operator Decree § 6 (MedProdBetrV). This regular TSI should be carried out
at least once a year. When this is done the functional and electrical safety in
accordance with VDE0751, including a visual and function inspection (in-
spection of the resource in operation) must be performed, as well as an in-
spection of possible measuring functions
In general the accessories and other devices that are combined with the
product must also be inspected for technical safety. It must be ensured when
doing this that only accessories authorised by the manufacturer and permit-
ted combinations of appliances are used. These TSIs may only be undertaken
by specialised electricians or persons instructed in electrotechnical matters
using a special measuring and test device, who have been trained specifically
by the manufacturer for the product and the permitted accessories and combi-
nations of appliances. The person performing the inspections may not accept
any instructions from the operator. Special test records must be written for
each of the TSIs. The product itself should be marked with a test sticker displa-
ying the next test date.
Page 96

Publicidad

Tabla de contenido
loading

Tabla de contenido