15. FUNCTION TEST

Before a nursing home bed is used for the first time, several function tests must be perfor-
been achieved and are made without interruption.
These are explained in more detail in the following chapter.

15.1. HANDSET "CARISMA 07 JPF" FUNCTION DESCRIPTION

Dio d e B
Dio d e A
Sicherheits-
Sic h er-
Safety key
schlüssel
he itssch lü s
l
Fig. 7
Caution: set the locking switch before you
start. (13.4)
Function 1 Backrest (BR) up / down
Function 2 Thighrest (TR) up / down
Function 3 Patient surface height
(HA) up / down
Arrow 4
Adjustment direction up
Arrow 5
Adjustment direction down
Function 6 Comfortable sitting position
Function 7 Legs up position
Diode (B) =
rently activated function key
Function key remains activated for approx. 10
sec. before moving automatically to locking
mode
Diode (A) = Initial fault tolerance check
If locking areas are released => LED lights up
green
If power units are in operation => LED lights up
red, as long as one of the action keys up/down
(key 4 or 5) is pressed.
Error function is displayed if
a) red LED is permanently on
b) when key is pressed, if no LED
display / beep occurs
c) Continuous beep
LED does not light up + beep when key is
pressed (end position has been reached)
-
Page 77