Where are the QR code and activation code located?
The codes are lasered onto the bottom section of the case on the mounting plate of the mini-computer (section
"Starting the app for the first time" – see Page 68).
Can different passwords be used for different users?
Yes, by assigning a different micro-SD card to each user. Then the app creates a separate database with a pass­
word on each micro-SD card.
Is it possible to display the total patient weight?
No, the app only displays the values separately for the two sides of the force measurement plate. The total value
can be calculated by adding the individual values. These weights are only approximate since the scale function of
the 3D L.A.S.A.R. Posture is not calibrated.
What happens if you forget the password?
The password cannot be reset or restored. Ottobock is also not able to open the database. Therefore, be sure to
store the password safely in order to avoid data loss.
Where is the password noted?
The password is selected once during setup. Subsequently it can be changed but not read.
Can I change my password?
The password can be changed in the settings (section "Settings" – see Page 75).
Is personal data safe against unauthorised access?
The data is stored in an encrypted database on the micro-SD card. Encryption is via the login to the app. PDF files
and Data Station files generated with the app are not encrypted. Therefore they need to be protected against unau­
thorised access and securely erased if necessary.
How can the app database be backed up?
By copying the contents of the micro-SD card to another storage medium.
Is it possible to save the data if the tablet is defective?
The data is stored on the micro-SD card. It can be removed and subsequently used in a replacement tablet,
provided it was not damaged. Taking backup copies of the SD card contents at regular intervals is recommended
so the data can be restored if necessary.
What can be done if the tablet cannot be turned on?
Check whether the charge level of the tablet is sufficient, and connect it to the battery charger if necessary. Hold
the On button long enough to turn on the device.
Can the tablet be used while it is charging?
CAUTION! To protect against leakage currents, the patient is not permitted to come into contact with the
tablet during charging, directly or indirectly (for instance through another person).
Setup
Can the device also be used outdoors?
Only if the ambient conditions permit (dry, constant lighting and solid/even surface) and safe use is assured.
Does the force measurement plate have to be centred in front of the case?
Yes, to achieve the best possible accuracy.
What is the best way to lay the cables from the case?
The camera cables should be kept out of the working area for the O&P professional as far as possible to avoid trip­
ping (section "Setup" – see Page 63). The cable between the case and force measurement plate should be laid as
straight as possible and has a detachable magnetic connection.
Using the product
Can the alignment guidelines be used in 3D mode?
The stored alignment guidelines only apply when 3D mode is deactivated.
Can the device be used for bench alignment without the L.A.S.A.R. Assembly?
No, since the weight of the prosthesis alone would be too low to work with sufficient accuracy.
Error messages
What can be done if the error message "Connection failed" is displayed repeatedly even though the
mini-computer has booted up?
Check the WiFi settings on the tablet to verify that WiFi is active and connected to the network named
"3D_L.A.S.A.R._20XXXXXXX". If necessary, disconnect the power plug from the mini-computer and connect it to
the outlet again to perform a restart.