>SERVICE II
▼▲ ME
CALIBRATION
Reset operating time
Maintenance interval
NOTE
NOTE
>CALIBRATION
▼▲ ME
Calibration factor 1,00
>SERVICE II
▼▲ ME
Calibration
RESET OPERATING TIME
Maintenance interval
>OPERATING TIME
▼▲ ME
000023h
Reset
34
Service II
The service II menu contains various functions for monitoring the
operation of the instrument:
Programming of the maintenance interval, operation (reset) internal
data/setting, e.g. water quantity and plant monitoring.
Use of the Service II menu
The functions in the Service II menu directly influence the operation
and monitoring functions of the instrument!
These tasks should only be carried out by trained and qualified
staff.
Calibration
Entering a calibration factor is not required under normal operating
conditions, as the instrument has been calibrated and tested at the
factory.
Should the measured value displayed nevertheless deviate from the
target value of a measured standard solution, you can adjust the
instrument with the calibration factor.
You can calculate the calibration factor as follows:
Target value
Calibration factor =
Actual value
The calibration factor can be between 0.9 and 1.1.
Should the calibration factor vary from 1.0 by more than 10%, send
the instrument back to the factory for inspection.
Select from the menu >
=>
=>
Service II
CALIBRATION
Enter the desired calibration factor
Confirm the selection by pressing "ENTER"
Reset operating time
After replacing a dosing pump or the measuring chamber holder, it is
possible to reset the current operating time to 0 hours.
In
, select
>BASIC PROGRAM
=>
=>
SERVICE II
RESET OPERATING TIME
Select "Reset" to reset the operating time
Press "ENTER" to confirm the selection
The operating time "000000h" appears on the display
BASIC PROGRAM
Service II