Limit value 2
terminals 6, 7, 8
M. point
switch-over
terminals 9, 10, 11
>MEASURING POINTS ▼▲ ME
1 Measuring Point
2 MEAS. POINTS INT.
*
2 Meas. Points ext.
AUX
terminals 12, 13
>FUNCTION AUX
▼▲ ME
CONTACT BEFOR ANALYSIS
*
*
Contact during analysis
Contact „Suction"
Time:
00m:10s
Function
LV2
Relay switches when the
limit value is exceeded at:
Limit value 2 or
measuring point 2
Please refer to the section entitled "Switch functions of the limit value
outputs LV1 und LV2" for more details and programming!
Measuring points 1 or 2 (measuring point switch-
over)
If the instrument is used for monitoring two measuring points, the
solenoid valves (individual valves or one 3/2-way control valve) of the
corresponding sampling line have to be connected to this output. The
terminals are strictly allocated to the measuring points.
Terminal 9 = Measuring point 1 , Terminal 10 = Measuring point 2
In
, select
>BASIC PROGRAM
=>
=>
PROGRAM VALUES
Select "2 Measuring points"
Press "ENTER" to confirm the selection
AUX (programmable function output)
The functioning of this volt-free relay output is programmable:
1. for reporting a current analysis and/or
2. for contact prior to an analysis, e.g. to operate with a
cooler or
3. to control a feed pump for sampling
In
, select
>BASIC PROGRAM
=>
=>
PROGRAM VALUES
Select the program step at which the AUX contact is to be
activated
Under "
" enter the contact duration in minutes (m) and
Time
seconds (s)
Press "ENTER" to confirm the entry
Description of the relay outputs
Contact
Action
Volt-free
Programmable:
change-
- Continuous contact
over
- Impulse (1-99 seconds/minutes)
- Interval (1-99 seconds/minutes)
contact
- Hysteresis (limit value is exceeded
once, twice, three times)
MEASURING POINTS
FUNCTION AUX
41