Note:
Setup
• Read and understand the manual for each optional accessory.
This manual does not contain cautionary information for optional
accessories.
• Exercise caution with the cables and arrange so that the patient does
not become entangled or bound.
Before use / during use
• Check the following after turning on the power:
- No smoke, abnormal odor, or abnormal noise is emitted.
- Press each button and verify that it operates.
- For functions that cause icons to light or flash, verify that the icons
light or flash (page 12).
- Measurement can be performed normally, and measurement error is
within the tolerance value.
• If the screen is not displayed normally, do not use the unit.
• When recycling or disposing of parts (including batteries) of the unit,
follow local government rules and regulations.
Cleaning
• For cleaning, see page 23.
Non-Invasive Blood Pressure (NIBP) measurement
• If the patient has acute inflammation, a pyogenic ailment, or an external
wound at the location where the cuff is to be wrapped, follow the
instructions of the doctor.
• Non-Invasive Blood Pressure measurement (NIBP) is performed by
compressing the upper arm.Some people may experience intense
pain, or transient spotting caused by subcutaneous hemorrhaging
(bruising) may appear. The spotting will disappear with time; however,
it may be appropriate to inform patients for whom this may be a
concern that spotting sometimes occurs, and if necessary, refrain from
measurement.
• To measure correctly, it is recommended that the patient relax and not
talk during measurement.
• To measure correctly, it is recommended that the patient rest quietly for
5 minutes before measurement.
8