Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 169
Check before Use
Before turning on the power
Before turning on the power, check for the following
■ External appearance
• The device or accessories are not deformed due to falling or other
impact.
• The device is not dirty.
• The device is not wet.
■ AC adapter
• The AC adapter is firmly connected to the connector on the device.
• There are no heavy objects lying on the AC adapter cable.
• The AC adapter cable is not damaged (core-wire exposure, breaks, etc.).
When turning on the power
When you turn on the power, check the
LCD display.
• When the [START/STOP] or [Power
ON/OFF] button is pressed to turn on
the power, the right screen appears.
After turning on the power
After turning on the power, check for the following
■ External appearance
• There is no smoke or odor coming from the device.
• The device is not making any unusual noises.
■ Buttons
• Press each button and check that it works.
■ Non-invasive blood pressure (NIBP)
• Make sure that a suitable OMRON GS CUFF is attached (one that fits
the circumference of the patient's arm).
• The cuff tube is firmly connected.
• The person checking the cuff should wrap the cuff around arm,
perform cuff measurement and check to see that blood pressure is in
the vicinity of normal measurements.
• While measurement is in progress, intentionally activate "Body
movement detection function" to halt discharge by bending the relevant
arm and move body and during this halt check that cuff pressure does
not drop.
24

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido