Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 169
List of Error Codes
■ Example: E2
SYSTEM
Error
Description
code
E9
Internal hardware error
Non-Invasive Blood Pressure (NIBP)
Error
Description
code
The cuff tube is not connected Firmly connect the cuff tube.
E1
Air is leaking from the cuff.
Did not inflate properly because
the arm or body moved during
measurement.
Moved body or arm during
measurement, or talked.
The cuff is not applied
correctly.
The sleeve is rolled up and is
E2
compressing the arm.
Measurement time has
exceeded specified time.
Specified time: 165 seconds
Points to check
Contact your OMRON retail
dealer or distributor.
Points to check
Replace with an OMRON GS
CUFF that does not leak.
Have the patient not move
the arm or body, and repeat
measurement.
Have the patient not
talk or move, and repeat
measurement.
Correctly apply the cuff.
Remove the garment and re-
wrap the cuff.
The measurement time
exceeds the expected time,
so the measurement was
ended in order to avoid patient
discomfort.
There is a possibility that
measurement is being repeated
over and over due to air leaking
from the cuff.
EN
29

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido