Measurement was not possible
Cause / solution
Check the patient by palpation or other method.
After checking the patient, check the error code and see "List of Error
Codes" (page 29) for Non Invasive Blood Pressure (NIBP) measurement.
Abnormal measurement value
Cause / solution
The causes below are possible. Check the patient by palpation and then
repeat measurement.
• Body movement (chills or other trembling)
• Arrhythmia.
• Noise in the cuff
- A nearby person touched the patient.
- Cardiac massage was being performed.
EN