Environmental Conditions
Temperature range: 5 to 40°C (41 to 104°F)
Operating
Humidity range: 15 to 85%RH (not condensed)
Conditions
Atmospheric pressure: 700 to 1060hPa
Temperature range: -20 to 60°C (-4 to 140°F)
Storage and
Humidity range: 10 to 95%RH (not condensed)
transportation
Atmospheric pressure: 500 to 1060hPa
Non-Invasive Blood Pressure (NIBP)
Measurement
Oscillometric
technology
Measurement
Dynamic Linear Deflation method
method
Pressure
0 to 300 mmHg
display range
Pressure display
Within ±3 mmHg
accuracy
NIBP
SYS 60 to 250 mmHg
measurement
DIA 40 to 200 mmHg
range
PULSE 40 to 200 /min
NIBP
Maximum mean error within ±5 mmHg
accuracy*
Maximum standard deviation within 8 mmHg
Pulse rate
Within ±5 % of reading
accuracy
EN1060-1:1995+A2:2009
Reference
EN1060-3:1997+A2:2009
Standard:
ISO81060-1:2007
* Comparison with auscultation method performed by a trained professional.
DIA determined by the auscultation method is "K5".
NOTE:
• Specifications may be changed without prior notice.
This blood pressure monitor fulfils the requirements of the EC directive
93/42/EEC (Medical Device Directive). It also conforms to the European
standard EN 1060, Non-invasive Sphygmomanometers Part 1: General
Requirements and Part 3: Additional Requirements for Electromechanical
Blood Pressure Measuring Systems.
EN
33