Check the insulation of the safety measuring cables. Discard the safety
measuring cables immediately if the insulation is damaged.
-
Check safety measuring lines for continuity. If the conductor in the safety
measuring line is interrupted, the safety measuring line must be quarantined
immediately.
-
Before - at the rotary switch a different function is selected, the safety
measuring lines must be disconnected from the measuring point.
-
Strong sources of interference in the vicinity of the BENNING CM 7 can
lead to unstable readings and measuring errors.
8.2 Voltagemeasuring
Donotexceedthemaximumpermittedvoltagewithrespectto
earthpotential!Electricaldanger!
The highest voltage which may be applied to the jacks,
-
COM socket
-
jack for V,
of the BENNING CM 7 against ground, amounts to 1000 V.
-
Use the rotary switch to select the required function (V AC) or (V DC) on
the BENNING CM 7.
-
The black safety measuring cable has to be contacted with the COM jack
on the BENNING CM 7.
-
The red safety measuring cable has to be connected to the jack for V,
on the BENNING CM 7.
-
Bring the safety measuring lines into contact with the measuring points,
read off measured value on the digital display on the BENNING CM 7.
See figure 2:
See figure 3:
8.3 Direct/alternatingcurrentmeasurement
8.3.1
Preparations for making measurements
Operate and store the BENNING CM 7 only at the specified storage and
operating temperatures conditions. Avoid continuous insulation.
Strong sources of interference in the vicinity of the BENNING CM 7 can lead to
unstable readings and measuring errors.
Do not apply any voltage to the output contacts of the
BENNINGCM7! Any possibly connected safety measuring
cableshavetoberemoved.
8.3.2
Direct/ alternating current measurement
-
Use the rotary switch to select the required function (A AC) or (A DC) on the
BENNING CM 7.
-
Operate opening lever , clamp single wire live conductor centrally by
means of the BENNING CM 7 current probe.
-
Read off the digital display unit .
See figure 4:
8.4 Resistancemeasuringandacousticcontinuitytesting
-
Use the rotary switch to select the required function (
BENNING CM 7.
-
The black safety measuring cable has to be contacted with the COM jack
on the BENNING CM 7.
The red safety measuring cable has to be connected to the jack for V,
on the BENNING CM 7.
-
Bring the safety measuring lines into contact with the measuring points,
read off measured value on the digital display on the BENNING CM 7.
-
If the conductor resistance between the COM jack and the jack for V,
30 , the fitted buzzer sounds on the BENNING CM 7.
See figure 5:
11/ 2006
Direct voltage measurement
Alternating voltage measurement
Direct/ alternating current measurement
Resistance measurements
B ENNING CM 7
) on the
4