Mettler Toledo SevenGo Duo Pro Instrucciones De Manejo página 16

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

13
Number of data sets in memory
14
User ID
15
Softkey
16
Softkey
17
Softkey
18
Sample ID
19
Sensor ID
20
DO OptiOx sensor cap lifetime icon
New sensor cap
Sensor cap needs to be re-
placed in less than 1 month
21
pH electrode condition criteria (only SG98)
Slope: 95-105%
Offset: ±(0-15)mV
Electrode is in good con-
di t ion
22
ISM
sensor connected
®
Stability criterion (only SG98)
Strict
24
Warning messages
25
Buffer groups or standards
14 Operating the pH/ORP/Ion/DO meter SG98 and DO meter SG9
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 12
Sensor cap needs to be re-
placed in less than 6
months
Sensor cap needs to be re-
placed in less than 2 weeks
Slope: 94-90%
Offset: ±(15-35)mV
Electrode needs cleaning
Medium
Sensor cap needs to be re-
placed in less than 3
months
Sensor cap needs to be re-
placed in less than 2 days
Slope: 89-85%
Offset: ±(>35)mV
Electrode is faulty
Fast
21.04.2011 11:45:51

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sevengo proSevengo optioxSg9Sg98

Tabla de contenido