− Liquids and foreign material (such as dust, metal etc) must not be allowed to enter
the device. If such material has entered into the units, it must be immediately checked
by a service technician, before it can be reused.
− electricity supply. Never connect the stimulation cables to an external power supply, as
there is a risk of electrocution.
− Do not apply stimulation near the area of an implant, such as cochlear implants,
pacemakers, skeletal anchorage or electric implants. This could cause an electrical
shock, burns, electrical interference or death.
− Never use the CEFAR TENS or the AC adaptor if it is damaged or open. There is a risk of
electric shock.
− Disconnect the AC adaptor immediately if there is abnormal heating or smell, or if
smoke comes from the AC adaptor or the device.
War n i n g!
Equipment malfunction - this warnings can prevent equipment malfunctions that result
in patient hazards
- Keep the unit away from such equipment and verify its performance before use.
- Do not use the CEFAR TENS within one metre of short wave or microwave devices as
this could alter the currents generated by the stimulator. If you are in any doubt as to
the use of the stimulator in close proximity to another medical device, seek advice from
the manufacturer of the latter or from your doctor.
- Exercise caution when using electrotherapy while the patient is connected to
monitoring equipment with electrodes attached to the body. Stimulation could disrupt
the signals sent to the monitoring equipment.
- Refer repair and maintenance to authorized persons. Persons are authorized after
training by a specialist trained and commissioned by the manufacturer.
- Inspect the CEFAR TENS and it's accessories for damage and loose connections at least
once a year. Damaged and worn parts must be immediately replaced with original
spare parts by authorized staff.
- Do not use the device if you are connected to a high-frequency surgical instrument as
this could cause skin irritation or burns under the electrodes.
C a ut i o n !
Patient hazard - these cautions need to be observed to avoid the risk of electrical
shock or other negative effects to the patient.
- Do not apply stimulation close to metal. Remove jewellery, piercings, belt buckles or
any other metallic product or device in the area of stimulation
- Be careful if the patient has sensitivity problems or is not able to communicate that he
e n