6 . 3 s e l e c t a pr e set P rog r am
Press the Program "P" button for the left or right channel you want to adjust. The program number is
blinking on the screen. Push on up/down arrows of the same channel to increase or decrease the
program number and confirm with "oK". If you want to use both channels, then select a program for
the other channel. Any combination of programs is possible except for Program 7, which must be
selected for both channels.
note: When selecting a program the amplitude must be set to 00.0 mA or 0.0V for the channel
concerned. For further information on the programs, see section PRESET PROGRAMS.
6 .4 A d j us t t i m e r se tti n gs if n eeded
Timer preset value can be either unlimited (--:--) or with a limited time . See program table
summary at the end of this manual. Press once the "timer" button, Channel 1 time value is blinking. To
adjust time press on adjust value buttons "up or down" and press "oK" to confirm
Press twice the timer button to select Channel 2 timer adjustment.