9 . t R o U B L e s H o o t I n G
9 . 1 e l e c t r o d e o r ca b le f a ilur e
The CEFAR TENS will display a pair of
electrodes and an arrow pointing to the
channel where the problem was detected.
The electrode error message can mean that:
• There are no electrodes connected to this channel.
• The electrodes are old, worn, and/or the contact is poor: try using new electrodes.
• The electrode cable is faulty: try connecting
it to another channel. If the problem persists, replace the cable.
If for any reason your device seems to present a malfunction other than those presented above,
please contact DJO Global customer service.
Problem
electrode failure
9 . 2 B at t er y L ev e l
Problem
Device battery low
Possible Cause
Poor electrode connection to
the cable
Poor electrode connection to
the skin
Possible Cause
During the stimulation the
device may turn out due to
discharged battery.
Solution
Check that the electrodes are
properly
connected to the cable.
Check if the electrodes are outdated,
worn and/or the contact is poor: try
with new electrodes
Solution
Stop the stimulation and recharge
the device.