C e F A R t e n s
stimulation is ineffective
stimulation only felt on one
electrode
stimulation on one channel
(side) only
Intermittent output
stimulation is not producing
the usual sensation
When an error is detected by the device (from anywhere in the device), the stimulation shall stop on
all channels and a specific screen will be displayed. If this screen is displayed, record the Error number
and call the DJO Customer Service.
Improper electrode placement
Unknown
Improper electrode placement
Electrodes
A. Worn or damaged
B. Improper placement
Intermittent Program in use
Settings and Electrodes
positioning
Reposition electrodes.
Electrodes must be a minimum
of 5 cm (2 inches) apart.
Contact clinician.
A. Reposition electrode.
Electrodes must be a
minimum of 5 cm (2 inches)
apart.
B. Replace electrodes.
A. Replace
B. Reposition electrode.
Electrodes must be a
minimum of 5 cm (2 inches)
apart.
Some programs will seem
intermittent. This is expected.
Refer to section 8.1 for a
description of the programs.
A. Check that all the settings
are correct and ensure the
electrodes are positioned
properly
B. Change the positioning of
the
electrodes slightly.