WARNING: DO NOT place dispersive electrode cables, active cords or
other electrosurgical accessory cables near cables leading to or from
cardiac or other patient monitoring equipment. DO NOT bundle cables or
cords with cables or cords from other electrical equipment as it may result
in interference.
WARNING: Capacitive Coupling. Wrapping electrosurgical instrument
active cords around metal objects can generate currents (capacitive
coupling) which may lead to electrical shock to the patient or user or fire.
WARNING: Alternate Site Burns or Shocks. DO NOT place active
portions of accessories in contact with grounded patients as a burn may
result if the unit is inadvertently activated when not in Standby Mode.
→ Activate ONLY when the active accessory is in view and
→ DO NOT use worn or ill-fitting active cords, dispersive
→ DO NOT leave the active accessory in the patient when not
WARNING: Electrical Shock Hazard. Follow the instructions for use
provided with the electrosurgical accessories for proper connection to the
electrosurgical generator. Prior to use, ensure all accessories are
properly connected to their receptacles and that no metal components of
the accessories are exposed as this may result in injury to the patient or
user.
Do not pull on the cable itself.
A defective Active Cord cannot be repaired. Discard and replace.
Rated accessory voltage of 3500V (7000Vp-p).
Contraindications:
Contraindications include those specific to any endoscopic procedure,
as well as those specific to endoscopic electrosurgery
Prior to Use:
Prior to clinical use, familiarize yourself with the device and read all the
Instructions for Use.
Inspect the package for shipping or handling damage. If damage
(breaks in insulation, exposed metal cable or wire, brittle
insulation) is evident, do not use this device and contact your local
STERIS Endoscopy Product Specialist and save the device and all
packaging.
00732676B
near or touching target tissue.
electrode cables or connector adapters as shock or burn can
occur.
in use.
Page 5 of 89