2.4 Installing the Filters and Tubing to the LEEP PRECISION Smoke Evacuator
2.4.1 ULPA Filter Installation
Tilt the LEEP PRECISION Smoke Evacuator forward and insert the large
ULPA Filter Cylinder with the Air Flow arrow pointing downward
(see Photo B) .
Photo C
2.4.3 Connecting the Tubing
There are two tubing hookup options:
1 . For procedures requiring close-proximity plume removal (i .e ., vaginal speculum)
2 . For procedures requiring open-area plume removal (i .e ., external lesions)
For procedures requiring close-proximity plume removal (i.e., vaginal speculum)
⅜ -inch Reducer (REF 6083) onto the port on the
Assemble the
disposable Pre-Filter (REF 6081) top with a slight twisting
motion . Attach one end of an appropriate length of
Evacuation Tubing (REF 6084) to the Reducer connector and
direct the other end to the patient and any appropriate device being used,
such as a vaginal speculum equipped with a
Smoke Evacuation Adapter (see Photo D) .
For procedures requiring open-area plume removal (i.e., external
lesions)
Assemble the sterile Disposable Evacuation Tubing (REF 6084) directly into the top of the Pre-Filter .
Position the opposite end over the site to be treated .
2.5 Connecting the LEEP PRECISION Integrated System
to the Wall Outlet
Make sure the ON/OFF Switch
Integrated System, located on the LEEP PRECISION Integration
Unit, is in the OFF position ("O"). Next, attach the power cord to
the receptacle
13
then into a hospital-grade grounded wall outlet to obtain power [refer
to (A) in Figure 1] .
2.4.2 Installing the Pre-Filter
Prepare the unit for use by inserting a clean, disposable Pre-Filter
(Single Patient Use) onto the ULPA Filter Cylinder (see Photo C) .
Be sure the device is firmly seated .
for the LEEP PRECISION
12
on the LEEP PRECISION Integration Unit and
⅜ -inch ID
13
8
Photo B
Photo D
12
14