Installing The Foot Pedal Switch On The Leep Precision Generator; Installing The Active And Dispersive Electrodes - Cooper Surgical LEEP PRECISION LP-10-120 Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 79

2.6 Installing the Foot Pedal Switch on the LEEP PRECISION Generator

Connect the Foot Pedal Switch to the socket
1

2.7 Installing the Active and Dispersive Electrodes

2.7.1 Placement of the Active Electrode
Connect the Active Electrode to socket
2.7.1.1 For Foot Switch-Operated Electrode Use (with Adapter)
A 4 mm Reusable Electrosurgical Adapter is supplied with the LEEP PRECISION
Integrated System (found in the LEEP PRECISION Generator box) . Use the Foot
Switch-Operated Electrode with the Adapter . The Electrode is plugged into the Adapter, which is then plugged
into socket
shown in Figure 2 .
7
2.7.2 Placement of the Dispersive Electrode or Patient Return Electrode
When using an electrosurgical system, it is very important that all of the current delivered to the
patient returns correctly to the LEEP PRECISION Generator via the Patient Return Electrode only .
Connect the Dispersive Electrode to socket
The patient must be positioned correctly on the operating table . The patient and operator must
not come in contact with any metal conductive surfaces .
The Patient Return Electrode must securely contact a vascular area close to the operating site . For a
gynecology procedure, the preferred site is the patient's thigh . The contact area must be clean, free of body
lotions, shaved, and massaged for good circulation . The Patient Return Electrode contact area must be
maximized and frequently checked for uniform contact during the procedure, especially if the patient has
moved or if liquids have come in contact with the Patient Return Electrode . The Patient Return Electrode
must NEVER be placed so as to allow the patient's heart to be in the pathway from the active electrode .
Power delivery to the operative site may be decreased appreciably if alternate pathways exist; for example,
through the metal operating table, crossed Active Electrode/Patient Return Electrode Cables, etc .
5
3
4
2
Figure 2
shown in Figure 2 .
7
shown in Figure 2 if foot control is desired .
5
. Refer to Figure 2 .
8
9
7
8

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Leep precision lp-10-220

Tabla de contenido