Operation
graphic presentation
8.6
Check therapy progress
Two different modes are available. Toggle between the two modes by pressing «Data»
or «Graph». In the data mode the set pressure value and the air flow value are digitally
displayed. The air flow value is displayed in real time equivalent to the respective paren-
chymal leakage.
Rounding Function
26 |
Preparation for use / operating instructions
Digital display of air flow values is rounded to zero ml / min
(0 ml / min) if the air flow measurement indicates an air flow
< 5.0 ml / min.
If the air flow measurement indicates an air flow ≥ 5.0 ml / min
and < 15 ml / min the digital display of air flow values is rounded
to ten ml / min (10 ml / min)
Air flow values are displayed
as follows:
– Air flow values between
0 and 1000 ml / min are
displayed in 10 ml steps
– Air flow values greater than
1000 ml / min are displayed
in 100 ml steps.