10 Changing pressure

WARNINGS
The settings may only be changed by the physician or on medical orders.

Change pressure

Gravity mode

32 |
Changing pressure
Important: Thopaz is running (Data- or Graph mode).
Press simultaneously
For patients who are to be treated by gravity drainage (= water
seal), the gravity mode can be activated. This mode corresponds
to a pressure of -0.8 kPa / -8 cmH
vate the gravity mode press the [
«OK».
1.1 Press [
] and [
simultaneously. The
pressure is indicated on
the display.
1.2 Set the desired pressure
with [
] and [
and confirm with «OK».
O / -8 mbar / -6 mmHg. To acti-