Technical Data - Ottobock 4R91 Instrucciones De Uso

Ocultar thumbs Ver también para 4R91:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 23
2.3 Function
The Tube Adapter is to be used in conjunction with the Tube Clamp Adapter
as an adjustable and length-adaptable connection element for Ottobock
Modular Lower Extremity Prostheses.
For maximum structural strength the Tube Clamp Adapter must be placed
proximally (as shown in the cover illustration). The clamping slot must always
be positioned anteriorly.

3 Technical data

2R57
220 g
Weight
34 mm
Diameter
System height
Min.system
77 mm
height
Max.system
282 mm 282 mm 472 mm 472 mm
height
Titanium INOX* Titanium INOX* Titanium Titanium INOX*
Material
Maximum
150 kg 150 kg 150 kg 150 kg 150 kg 150 kg 150 kg
patient weight
* stainless steel
4 Handling
4.1 Connecting the internal adapter with tube to the tube clamp ad-
apter
The Tube Adapter must be shortened as necessary based on the amputee's
measurements. This must be done correctly, as noted below, to avoid dis-
torting the tube. Due to the close tolerances of quality Ottobock adapters,
a distorted tube will not fit easily into the Clamp Adapter.
Attention!
Do not clamp the tube in a vise, or it may become permanently
disorted!
The cut surface must be perpendicular to the long axis of
the tube (ill. 1). Use of the 719R2 Tube Cutter or equivalent is
recommended.
10
2R76
260 g
34 mm
77 mm
2R58
2R77
330 g
370 g
34 mm
34 mm
77 mm
77 mm
4R82
4R82=P
95 g
90 g
34 mm
34 mm
33 mm
-12 mm
4R91
140 g
34 mm
33 mm

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4r822r572r764r82-p2r582r77

Tabla de contenido