Problem: Hose
Solution: Replace the hose between the vacuum con
nector of the prosthetic socket and the suction connector
of the prosthetic knee joint.
= The vacuum does not decrease within 20 seconds.
= The vacuum decreases within 20 seconds.
9 Disposal
In some jurisdictions it is not permissible to dispose of the product with unsorted household
waste. Improper disposal can be harmful to health and the environment. Observe the information
provided by the responsible authorities in your country regarding return, collection and disposal
procedures.
10 Legal information
All legal conditions are subject to the respective national laws of the country of use and may vary
accordingly.
10.1 Liability
The manufacturer will only assume liability if the product is used in accordance with the descrip
tions and instructions provided in this document. The manufacturer will not assume liability for
damage caused by disregarding the information in this document, particularly due to improper
use or unauthorised modification of the product.
10.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745 on medical devices. The CE
declaration of conformity can be downloaded from the manufacturer's website.
11 Technical data
Article number
Weight [g]
Proximal system height up to the alignment reference point
[mm]
Distal system height [mm]
Proximal connection
Distal connection
Max. flexion angle [°]
Problem: Valves or vacuum pump
Solution: Remove the valves of the 3R60 Vacuum;
clean or replace them. Connect the vacuum gauge to
the suction connector of the prosthetic knee joint.
Flex and extend the prosthetic knee joint until a vacu
um of at least 300 mbar is reached. Check whether
the vacuum level is maintained.
Problem: Vacuum pump
Solution: Please contact the manufacturer's local
service representative and send in the defective pros
thetic knee joint.