Warning instructions: When cleaning the reusable
ear funnels, special attention is required. A tempera-
ture of 134°C in the autoclave is not to be exceeded.
When switched on for a long time the parts around the
bulb can become very hot. As the otoscope is not in-
tended for use during operational procedures, simple
cleaning and or disinfection with an alcohol-based
surface disinfecting agent is sufficient.
KaWe otoscopes should not be used in rooms where
the temperature may exceed 35° C. KaWe otoscopes
are subject to special safety measures regarding elec-
tromagnetic compatibility (EMC). The operation of the
KaWe otoscope can be influenced by portable and
mobile HF communication devices.
Limitation of reprocessing: Frequent reprocessing
influences the instruments only minimally. The work-
ing life of the instruments usually depends on the wear
and tear and damage by use. It is recommended, that
the ear funnels be re-prepared as soon as possible after
their use on a patient.
Instructions
Workstation: Remove dirt accumulation on surfaces
with a paper/disposable cloth.
Storage and transport: No special requirements
Cleaning preparation: No special requirements
Manual cleaning: The exterior surface of the veteri-
nary otoscope can be cleaned with a damp, soft, lint-
free cloth. Never allow the otoscope to come into con-
tact with fluids and ensure that liquids do not
penetrate the housing. Do not use abrasive cleaning
agents! The unit cannot be sterilized. The magnifying
glass is to only be cleaned with glass cleaner.
Automatic cleaning: No automatic cleaning possi-
ble/required.
Disinfection: Only the reusable ear funnels need to
be disinfected with (for example) helipur. Follow the
concentrations and reaction times given by the manu-
facturer. The veterinary otoscope can be disinfected by
wiping it off with a cloth dampened with alcohol.
Sterilization: Reusable funnels must be sterilized in
an autoclave at a temperature of 134°C for a minimum
of 3 minutes.
Additional information: If different kinds of instru-
ments are being sterilized in the autoclave at the same
time, the maximum load given by the manufacturer
must not be exceeded.
Maintenance: Defective parts must be removed and
replaced. If the device does not function properly, fol-
low these steps: 1. Replace batteries/rechargeable
batteries and check their functionality; 2. Replace the
bulb and check its functionality; 3. Send the unit in for
repair.
Check and performance test: Check all movable
parts such as the frames of the magnifying glasses and
plug-in connections to ensure that they all move
smoothly. Turn on the unit. For all units: Visually check
all parts for damage and wear.
Packaging: Individually packed in a cloth bag or box.
Storage: No special requirements.
8