12
Re-use
The Solstice Comfort wheelchairs are designed for re-use.
They must be cleaned and disinfected prior to re-use.
If disinfection is required, only use suitable disinfectants
in accordance with the scope of the wheelchair.
Note
If you are uncertain or have questions, please contact
your Days Healthcare specialised dealer (see dealer stamp
on the back cover).
12.1
Drive wheel – Removal and installa-
tion
12.1.1 Removing the drive wheel
1. Press and hold the interlock (A) of the quick-release
axle
2. Pull the drive wheel together with the wheel axle out
of the mounting socket (B).
12.1.2 Installing the drive wheel
Risk of accidents!
Check the wheel locks after every wheel change to en-
sure perfect function and readjust them, if necessary.
A
B
Solstice Comfort – 15.09.2010
A
1. Press and hold the interlock (A) of the quick-release
axle.
D
B
C
2. Push the quick-release axle (B) completely into the
mounting socket (C) and release the interlock (A).
Ensure that the retaining balls (D) at the end of the
quick-release axle protrude beyond the mounting
socket (C).
Risk of accidents!
When the retaining balls (D) do not protrude out of the
quick-release axle, the affected drive wheel is not secured
and can fall off when the wheelchair is moved.
12.2
Drive wheels – Replacing the tyre
Note
• The drive wheels (A) are equipped with puncture-
proof (tubeless) PU tyres. If a tyre is heavily damaged
and the driving characteristics of the wheelchair are
impaired, the affected tyre must be replaced.
• In this case, please contact your Days Healthcare spe-
cialised dealer (see dealer stamp on back cover).
A