The Aaron 940 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the
output power of the communications equipment.
Rated maximum output
power of transmitter
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
The Aaron 940™ is intended for use in the electromagnetic environment listed below. The customer or the
user of the Aaron 940™ should assure that is used in such an environment.
Emissions test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
User's Guide • Aaron 940 ™
Recommended separation distances between portable and mobile RF
communications equipment and the Aaron 940™.
Aaron 940™
Separation distance according to frequency of transmitter in metres (m)
150 kHz to 80 MHz
d = 1.2 P
W
0.12
0.01
0.38
0.1
1.2
1
3.8
10
12
100
Guidance and manufacturer's declaration – electromagnetic emissions
Group 2
Class A
Class A
Complies
Aaron 940™
can help prevent electromagnetic
as recommended below, according to the maximum
80 MHz to 800 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
Compliance
Electromagnetic environment - guidance
Aaron 940™
The
electromagnetic energy in order
to perform its intended function.
Nearby electronic equipment may
be affected.
Aaron 940™
The
use in all establishments other
than domestic and those directly
connected to the public low-
voltage power supply network that
supplies buildings used in
domestic purposes.
800 MHz to 2.5 GHz
d = 2.3 P
0.23
0.73
2.3
7.3
23
must emit
is suitable for
11