Bovie AARON 940 Guia Do Utilizador página 13

Ocultar thumbs Ver también para AARON 940:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 83
The Aaron 940™ is intended for use in the electromagnetic environment listed below. The customer or the
user of the Aaron 940 should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE U t is the a.c. mains voltage prior to application of the test level.
12
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5 % U t
(>95 % dip in U
)
t
for 0.5 cycle
40 % U t
(60 % dip in U
)
t
for 5 cycles
70 % U t
(30 % dip in U
)
t
for 25 cycles
<5 % U t
(>95 % dip in U
)
t
for 5 sec
3 A/m
Compliance level
Electromagnetic environment - guidance
Floors should be wood, concrete
or ceramic tile. If floors are
±6 kV contact
covered with synthetic material,
±8 kV air
the relative humidity should be at
least 30%.
±2 kV for power
Mains power quality should be that
supply lines
of a typical commercial or hospital
environment.
Not Applicable
±1 kV differential
Mains power quality should be that
mode
of a typical commercial or hospital
±2 kV common
environment.
<5 % U t
(>95 % dip in U
)
t
for 0.5 cycle
Mains power quality should be that
of a typical commercial or hospital
40 % U t
(60 % dip in U
)
environment. If the user of the
t
Aaron 940 requires continued
for 5 cycles
operation during power mains
70 % U t
interruptions, it is recommended
(30 % dip in U
)
t
that the Aaron 940 be powered
for 25 cycles
from an uninterruptible power
supply or a battery.
<5 % U t
(>95 % dip in U
)
t
for 5 sec
Power frequency magnetic fields
should be at levels characteristic
3 A/m
of a typical location in a typical
commercial or hospital
environment.
Bovie Medical Corporation

Publicidad

Tabla de contenido
loading

Tabla de contenido