Warranty And Repair; Troubleshooting; Transport And Storage; Output Power Characteristics - Bovie AARON 940 Guia Do Utilizador

Ocultar thumbs Ver también para AARON 940:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 83

WARRANTY AND REPAIR

The Aaron 940™ is covered under warranty for a period of two years. The handpiece is covered under warranty
for a period of one year or 25 steam autoclave cycles, whichever comes first. The warranty becomes null and
void if damage occurs from incorrect handling or misuse of the product.
For warranty and repair work, please contact Bovie® and obtain a Return Goods Authorization number (RGA).
Place the number so that it can be seen on the exterior of the package and ship directly to Bovie. A return
without an RGA may not be accepted.

TROUBLESHOOTING

The Aaron 940™ has been designed and manufactured with the utmost safety in mind. The unit is equipped to
automatically detect a malfunction. The following table list error codes, their meaning and
recommended actions to be taken to resolve the error.
Error Code
Description of Error
E1
Activation upon power up
E2
DC Supply Over Voltage
Detection
E3
Pulse Width
E4
Delta Error
E5
Multiple Errors
E6
Temperature Error
If problems persist, the unit should be taken out of service and the manufacturer should be notified. For
technical support or return authorization phone +1-800-537-2790.

TRANSPORT AND STORAGE

Ambient temperature range: -40˚C to 70˚C
Relative humidity: 10% to 100%, including condensation
Atmospheric pressure: 50kPa to 106kPa.

OUTPUT POWER CHARACTERISTICS

Maximum Output for Bipolar and Monopolar
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
Mode
Output Power
40 W @ 1000 Ω
Coagulation
40 W @ 200 Ω
Bipolar
14
Recommended Action
• Check handpiece for activation.
• Check footswitch for activation; once the activation is
halted the unit will resolve the error. If the error
persists the handpiece could be malfunctioning and
may need to be replaced.
• Switch unit off and on again.
• Make sure unit is connected to correct power source
for the unit.
• Switch unit off and on again.
• Switch unit off and on again.
• Switch unit off and on again.
• Switch unit off. Allow unit to cool. Switch unit on.
Output Frequency
Repetition Rate
550 kHz ± 50 kHz
21 kHz ± 10%
550 kHz ± 50 kHz
21 kHz ± 10%
Vp-p max
8.0 kV
2.5 kV
Bovie Medical Corporation

Publicidad

Tabla de contenido
loading