EN
12.0 CALIBRATION
Occasionally the scale should be verified whether it is
weighing correctly by measuring to a known mass.
Zero the scale. Place the mass on the centre of the platform
and note the reading. Calibrate the scale, if necessary.
NOTE: Before calibrating your should make sure you have the correct
weights. The weights should be known to an accuracy that is appropriate
for the balance being calibrated, for example, OIML Class M1 type or
ASTM E617 Class 4. If you do not have the correct weights do not
attempt calibration.
PROCEDURE
While in the normal weighing mode, press and hold the
[Tare/Zero] key for 4 seconds.
The display will show "CAL" along with the last selected unit.
The unit can be changed by using the [Unit] key to calibrate
in kg or lb.
Press the [Print/Hold] key.
where xx is the Calibration weight which is user-selectable.
Use the [Tare/Zero] key to change the flashing digit and the
[Print/Hold] key to move to the next digit.
Press the [Unit] key to confirm the calibration weight. The
digit stops flashing.
Note: If the selected mass is less than 10% of the capacity of
the scale, an error message "CALEr" will be displayed and the
scale will return to zero. Repeat the process correctly.
© Adam Equipment Company 2014
The display will show "L xx"
14