EN
Place the correct calibration mass as selected by the user at
the centre of the pan.
Press the [Unit]. The display will return to weighing mode.
Note: If the mass loaded is more than ±20% of the factory
calibration reference then an error message "CALEr" will be
displayed and the scale will return to weighing without
calibration being saved. Repeat the process correctly.
Remove the weight.
Verify the scale is calibrated correctly. Repeat the process, if
necessary.
13.0 PARAMETER SETTING
13.1 USER PARAMETERS
The scale can be set as desired by the user to control the weighing
operation. See section 18.0 for the complete list of parameters.
Switch off the scale.
Hold the [Tare/Zero] key and then press the [On/Off] key
momentarily. Release the [Tare/Zero] key. The display
shows the first parameter - auto power off.
To exit the parameter setting at any time, press the
[Print/Hold] key.
To scroll through the user parameters, press the [Unit] key
(which will advance to the next parameter).
To return to normal weighing, turn the scale off and back to
on again or press the [Print/Hold] key.