Probe Attenuation Level And Type (Od-411/Od-415); Trigger; Trigger Type - Promax OD-405 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5.6.9 Probe attenuation level and type (OD-411/OD-415)

On the OD-411/OD-415 models, the probe can be set to either voltage
or current. A signal probe has an attenuation switch to lower the original DUT
signal level to the oscilloscope input range, if necessary. The probe
attenuation selection adjusts the vertical scale so that the voltage or current
level on the display reflects the real value, not the attenuated level.
Procedure
1.
Press the CH1 or CH2 key.
2.
Press F4 repeatedly to select voltage or current probes.
3.
Use the variable knob to edit the voltage or current attenuation.
4.
The voltage/current scale in the channel indicator changes
accordingly. There is no change in the waveform shape.
Range
Note: The attenuation factor adds no influence on the real signal; it only
changes the voltage/current scale on the display.
5.7

Trigger

The trigger function configures the conditions by which the oscilloscope
captures the incoming signals.

5.7.1 Trigger type

Edge
Video
Pulse
11-2011
0.1X~2000X (1-2-5 steps)
Triggers when the signals crosses an amplitude threshold in
either positive or negative slope.
Extracts a sync pulse from a video format signal and
triggers on a specific line or field.
Triggers when the pulse width of the signal matches the
trigger settings.
USER'S MANUAL. OD-405/410B/411/415