5.7.3 Configuring Holdoff (OD-411/OD-415)

The Holdoff function is only available on the OD-411/OD-415 models. It
defines the waiting period before the scope starts triggering again after the
trigger point. The holdoff function is especially useful for waveforms with two
or more repetitive frequencies or periods that can be triggered.
Panel operation
1.
Press the Trigger menu key twice.
2.
To set the Holdoff time, use the Variable knob. The resolution
depends on the horizontal scale.
Range 40ns~2.5s
Pressing Set to Minimum sets the Holdoff time to the minimum, 40ns.
Note: The holdoff function is automatically disabled when the waveform
update mode is in Roll mode.

5.7.4 Configuring the edge trigger

Procedure
1.
Press the trigger menu key.
2.
Press Type, the F1 function key repeatedly to select edge trigger.
3.
Press Source repeatedly to select the trigger source.
Range
4.
Press Mode repeatedly to select Auto or Normal trigger mode. To
select the single trigger mode, press the Single key.
Range
5.
Press Slope/coupling to enter into the trigger slope and coupling
selection menu.
Page 71
Channel 1,2, Line, Ext
Auto, Normal