Operation

This section describes the following operational
procedures and functions of the MiniOX
Oxygen Monitor:

Calibration

Two-Point Linearity Check
Setting the Alarms
Alarm Conditions
Silencing an Alarm
Test Functions
Installing the Sensor in a Breathing Circuit
Mounting the Instrument with a Pole Bracket
Calibration
The MiniOX
3000 Oxygen Monitor requires
®
calibration:
daily, while in operation
when the operating environment changes
after the monitor is turned ON
EXPOSE SENSOR
1
TO ROOM AIR
2
PRESS 21%
3
CALIBRATE
3000
®
PRESS I/0
Calibrating the MiniOX
®
711449 (Rev.3) 02/2016
if the sensor is disconnected and reconnected
to the instrument
Recommended practice of calibration is against
room air (defined as oxygen concentration of 20.8%).
When a more precise measurement is desired,
repeat calibration using 100% oxygen concentration.
Changes in pressure, humidity or temperature may
affect calibration accuracy (See Appendix B, Effects
of Pressure, Humidity, and Temperature).
Do not handle the sensor unnecessarily
during calibration or use. Body heat can
cause the sensor's thermistor to change
disproportionately to the change in gas
sample temperature at the sensing electrode.
This can produce some error until thermal
equilibrium is restored.
To Calibrate In Room Air (Figure 3-1):
1. Expose sensor to room air.
2. Press I/0 to turn ON the instrument.
"CAL" flashes in the display.
3. Press 21%. The following appears on the
display:
CALIBRATION COMPLETE AND
Figure 3-1
3000 Oxygen Monitor in Room Air
Section 3, Operation

CAUTION

4
PRESS
UNLOCK
AFTER 20 SECONDS
20.8% + 2% DISPLAYED
5
3-1