To Calibrate At 100% O (Figure 3-2); 3000 Oxygen Monitor At 100% O - Ohio Medical Corporation MiniOX 3000 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

MiniOX
3000 Oxygen Monitor
®
"CAL"
"21% CAL"
4. Press
. The following appears:
"CAL"
"21% CAL"
a 10-segment bar graph that "counts down"
two seconds per bar for 20 seconds
5.
After 20 seconds, the calibration process is
complete the device:
displays 20.8% ±2%O
proceeds to the monitoring mode
displays the current oxygen concentration
as %O
2
To Calibrate at 100% O
Note: Prior to calibrating at 100% O
3000 Oxygen Monitor must first be calibrated
in room air.
1. Calibrate in room air (See Section 3, To
Calibrate In Room Air).
CALIBRATE IN ROOM
1
AIR
EXPOSE SENSOR TO
2
100% OXYGEN
PRESS 100%
3
CALIBRATE
Calibrating the MiniOX
3-2
(18.8% to 22.8%)
2
(Figure 3-2):
2
, the MiniOX
®
2
Figure 3-2

3000 Oxygen Monitor at 100% O

®
711449 (Rev.3) 02/2016
2. Expose the sensor to 100% oxygen and allow
the readings to stabilize prior to initiating the
calibration.
"CAL" flashes in the display.
3. Press 100%. The following appears on the
display:
"CAL"
"100% CAL"
4. Press
The following appears on the display:
"CAL"
"100% CAL"
a 10-segment bar graph that "counts down"
two seconds per bar for 20 seconds
5.
After 20 seconds, the calibration process is
complete; the device:
displays 100.0% +0/-2% (98% to 100%)
proceeds to the monitoring mode
displays the current oxygen concentration
as %O
2
PRESS
UNLOCK
AFTER 20 SECONDS
CALIBRATION COMPLETE
100.0% +0/-2%
AND
DISPLAYED
Section 3, Operation
4
5
2