DEVICE SPECIFICATIONS

Maintenance

At least once a year, the CP should be inspected for
visible damage and should be charged and powered on
to confirm the Log-In screen is still accessible . Significant
physical damage or an inability to power on the device
should be reported to the manufacturer and the device
should not be used .

Handling and Disposal

Cleaning: The CP can be wiped with a cloth lightly
moistened with water . No other cleaning agents should
be used .
Replacement: If the CP is lost or not working, contact
Axonics .
Disposal: Do not incinerate the CP as the battery may
explode . If the CP is no longer needed, contact Axonics
to return the device .
10 10

Wireless Communication

Radiofrequency telemetry
Model: 2501
IC: 20225-C
FCC ID: 2AEEGC
Quality of Wireless Service:
This device operates in the 401-406 MHz frequency
and the maximum effective radiated power is
below the limit of 25 µW ERP/EIRP as specified in
EU: EN ETSI 301-839 and EN ETSI 302-537 and USA:
FCC 47 CFR Part 95; Subpart I . The CP has to be
within 1 meter from the Stimulator for successful
communication .
Wireless Security:
Any CP can communicate with a Stimulator .
Additional mechanisms exist to ensure the integrity
of radio data .