Viewing the Status of a Stimulator
E0
E1
E2
E3
1
2
Therapy History Tab ( 1 ): This tab shows the stimulation
settings saved to the Stimulator in the previous 4 programming
sessions . For each set of stimulation settings, additional
information is provided including the time period the settings
were used and notes entered during the programming session .
The usage information includes:
Daily Usage: Provides a percentage that reflects the daily use,
or on time, of stimulation .
Weekly Adjustments: Displays a numeric value of weekly
adjustments based on adjusting stimulation levels by Patient
Remote . Momentary adjustments will not be recorded .
Average Amplitude: Displays average stimulation amplitude
based on usage and adjustments .
Threshold Detail Tab ( 2 ) (Tined Lead Trial only): This tab
summarizes the stimulation threshold data previously captured .
For each electrode, the tab shows the motor response, sensory
response, and threshold stimulation amplitude recorded during
the tined lead implantation procedure or during a programming
session . The color around the threshold amplitude indicates the
qualitative assessment of the electrode placement
(Red = "Poor", Yellow = "OK", Green = "Good") .