Controlling Stimulation Amplitude
2
3
1
8
3
7
22 22
2
5
4
6
9
7
Automatic Amplitude Adjustment
In the automatic amplitude adjustment mode the stimulation
amplitude automatically increases from 0mA to the amplitude
value set by the user . The amplitude increase can be stopped, or
stimulation can be turned off, at any time .
To use the automatic amplitude adjustment feature:
Use the slider bar ( 4 ) and up and down arrows ( 3 ) to set a
•
desired stimulation amplitude .
Press "Auto" ( 7 ) . "Auto" will be dark gray when automatic
•
amplitude adjustment is active .
Press the stimulation button ( 8 ) to turn on stimulation .
•
The stimulation amplitude will start increasing from 0mA
•
at a rate of 0 .2mA per second, and the output stimulation
amplitude will be visible in the digital display ( 2 ) at the top
of the stimulation bar . A red bar on the stimulation bar will
represent the stimulation amplitude as it increases ( 9 ) .
Note: The stimulation bar and up and down arrows are not accessible to
adjust stimulation .
To stop the increase in stimulation amplitude while still
•
delivering stimulation, press "Auto" ( 7 ) .
To stop stimulation altogether, press the stimulation button
•
( 8 ) .