Foramen Needle Test Simulation
8
5
6
1
9
4
7
2
3
To deliver stimulation, set the desired stimulation amplitude
on the Stimulation Bar ( 6 ) (see Controlling Stimulation
Amplitude for more information) . Next press the stimulation
button ( 7 ) to turn on stimulation . The stimulation button will
light up when stimulation is on .
Only stimulation amplitude is adjustable during foramen
needle stimulation . The stimulation frequency is 14 Hz, and
the pulse width is 210μs. The stimulation frequency can be
changed by exiting Lead Placement and going into CP Settings .
Note: The stimulation cable can be disconnected from the foramen needle
without causing stimulation to be turned off . This allows stimulation to be
delivered on demand by tapping the mini-clip portion of the cable against the
appropriate area of the needle .
Note: If test stimulation cannot be stopped using the stimulation button,
disconnect the cables from the CP .
The area around the tip of the foramen needle graphic
( 5 ) will be colored to reflect the proximity of the needle
to the sacral nerve . This information is used to evaluate
the proximity of the needle to the sacral nerve once the
appropriate muscle responses are observed (anal bellows
and big toe flexion) .