9.7. How to adjust the basic parameters of the movement?
Beginning
To stop the unit
To display the extension or flexion limit of the
movement
To change the limit if necessary
To confirm the new value, press another key
or wait more than 3 seconds for automatic
confirmation.
To display pause in flexion limit
or in extension limit
To change the pause if necessary
To confirm the new value press another key or
wait more than 3 seconds. The display shows
the selected mode
9.8. How to use the WARM UP key?
Warm Up rules.
The device Kinetec Spectra XL™ starts at 70% of the full ROM, increasing 5% of the
range each full cycle until the pre-set ROM is reached.
Beginning
To stop the unit
To select Warm Up mode
To start the movement
Note:
- The parameters should only be modified while the unit is stopped.
- Pauses and ByPass mode are not available during warm up cycles.
- The Warm Up cycles are only performed when the movement is first started.
- To start a new Warm Up session, press the "Warm Up" key twice.
- The calculation mode used enables the pre-set ROM to be reached in approximately seven full cycles.
Example:
For a Warm Up treatment with a pre-set ROM from 0° to 100°.
The first cycle starts with 15° to 85° to 15° and the values increase 5% each cycle.
Kinetec Spectra XL™
Keys to
Display
press
KINETEC
30 STOP
KINETEC
30
EXT
KINETEC
Or
40 FLEX
KINETEC
40 FLEX
Or
KINETEC
SPEED 2
KINETEC
FORCE _ _ _ _ _ _
KINETEC
TIMER 00H00MIN
KINETEC
PAUSE FLEX
KINETEC
PAUSE EXT
KINETEC
PAUSE EXT
Or
KINETEC
40 STOP
Keys to
Display
press
KINETEC
40 STOP
WARM UP
Please Wait
WARM UP
40 STOP
WARM UP
40 RUN
EN
Check if the hand control is not
45 70
locked (See page 7).
45 70
The value blinks.
45
70
The new value blinks.
45
110
While the value blinks press
the
or
to change if necessary.
0S
Successive presses on this key selects
the pause at the extension or flexion
limit.
0S
The new pause value blinks.
15S
The unit is ready to start with the new
45 110
parameters.
100°
50°
0°
Check if the hand control is not
45 110
locked (See page 7).
To change the movement value if
45 110
necessary (see page 9).
The angle display changes with the
50
110
current movement.
Remarks
key
Remarks
9/20