12. Setting up the patient
See page 4, "Installing the Device" chapter, for the positioning conditions.
Place the unit in a position that will be comfortable for the patient.
- Measure the length of the patient's femur in cm or inches (L); adjust the thigh support to this
measurement using knobs (5).
- Install the patient on the machine.
- Bring the foot plate (3) into contact with the patient's foot, then tighten both buttons (6).
- Adjust the plantar flexion (40°) or the dorsal flexion (30°) of the foot, with the knobs (7).
- Adjust the internal (30°) or external (30°) flexion of the foot, with the knob (27).
and the axis of the patient's knee (30) with the axis of articulation (31) of the unit.
7
13. Options
Trolley for all CPM
Part number to order:
4655001053
Box of USB keys and software
Part number to order:
4670025632 (5 USB keys + software)
4670025640 (20 USB keys + software)
Kinetec Spectra XL™
Adjust the axis of the patient's hip (28) with the
"THEORETICAL" axis of rotation (29) of the unit,
3
7
27
Cart for bed use
Part number to order:
4665003297
CAUTION
6
31
EN
30
5
Seat Adaptor
Part number to order:
4670024098
L
29
28
14/20