Masimo Iris Instrucciones De Uso página 3

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 16
Masimo®
Iris™ Adapter
Reusable
Prior to using this Adapter, the user should read and understand the Operator's Manual for the device,
monitor, cables, software/systems and this Directions for Use.
INTENDED USE
The Iris Adapter (Adapter) is intended to be used as a communication means between 3rd party medical devices and Masimo
Gateway via Iris connectivity ports.
DESCRIPTION
The Iris Adapter is a Masimo proprietary adapter, wired or wireless, which establishes a data communication between
3rd party medical devices (POCs) with Masimo Gateway, via Iris connectivity ports.
The Iris Adapter allows a variety of standalone devices (such as: IV pumps, ventilators, beds, and other patient monitors)
to connect to Root®, where the "3rd party" device data can be passed through Root to Masimo systems, such as Patient
SafetyNet™ or Iris Gateway™ and send the data to other systems.
For operation, the Iris Adapter requires Root and the Iris Gateway or Patient SafetyNet Supplemental Alarm System Series
5.0.0.0 (or higher) and compatible 3rd party device.
WARNINGS, CAUTIONS AND NOTES
• For complete safety information, consult the operator's manual for Root, and Iris Gateway or the Masimo
Patient SafetyNet Supplemental Alarm System Series 5.0.0.0 (or higher).
• Warning: Disposal of product - Comply with local laws in the disposal of the device and/or its accessories.
• Warning: Do not place containers with liquids on or near the Iris Adapter. Liquids spilled on the Iris
Adapter may cause it to perform inaccurately or fail.
• Caution: To prevent damage, do not soak or submerge the Iris Adapter in any cleaning solution or
attempt to sterilize by autoclave, irradiation, steam, gas, ethylene oxide, or any other method. This will
seriously damage the Iris Adapter.
• Note: Check the compatibility of the Iris adapter with the 3rd party device before use. The Iris adapter is
designed to work properly only with compatible 3rd party devices.
INSTRUCTIONS
A) Using the Iris Adapter
To connect a standalone device using Iris Adapter
1. Connect the "standalone" or "3rd party" device to to the Iris Adapter. See Fig. 1 (A).
2. Use an Ethernet cable to connect end Fig. 1 (B) of the Iris Adapter to any of the Four (4) Iris connectivity
ports on Root.
To confirm that the standalone device has successfully connected, see:
1. The the "standalone" or "3rd party" device's User Manual.
2. The Operator's Manual for Root.
3. The Operator's Manual for Patient SafetyNet Supplemental Alarm System Series 5.0.0.0 or higher or Iris
Gateway.
CLEANING
Wipe down the outer surface using any of the following solvents: Cidex Plus (3.4% glutaraldehyde), 10% bleach
solution, or 70% isopropyl alcohol solution. Do not allow liquids to enter the interior.
CAUTION:
• Do not immerse the connector in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
D IREC TIO N S FO R US E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
3
en
Non-Sterile
7785C-eIFU-1017

Publicidad

Tabla de contenido
loading

Productos relacionados para Masimo Iris

Tabla de contenido