COM4 Connections (POWERCELL Version Only)
The COM4 port on the POWERCELL version main board (Figure 2-26) provides an electrically
isolated RS-232 connection. Connections to this terminal should be made as shown in Figure
2-93.
2.4.6.
Wiring Connections for Options
IND570 terminal options that require external connections include the following:
Options are installed on the main PCB in the locations indicated in Figure 2-94.
30205319
|
11
|
07/2019
Ethernet TCP/IP
COM2/COM3 Serial Ports
5 In/8 Out Discrete I/O (Relay)
COM2/COM3 Serial Ports with 2 In/5 Out Discrete I/O (Relay)
Analog Output
ControlNet
DeviceNet
EtherNet/IP – Modbus TCP
PROFIBUS (Harsh Enclosure)
PROFIBUS (Panel-Mount Enclosure)
PROFINET
NOTES:
1.
USE ONLY SHIELDED CABLE.
2.
THE PORT IS ISOLATED SO ONLY THE GROUND
ON THIS TERMINAL SHOULD BE REFERENCED
TO THE CONNECTED DEVICE.
3.
MAXIMUM CABLE LENGTH: 50 FEET (15
METERS).
4.
WIRE SIZE: 18 AWG (.823 mm
TERMINAL
TXD
RXD
GND
Figure 2-93: COM4 Wiring
METTLER TOLEDO IND570 Terminal Installation Manual
2
) MAXIMUM
2
24 AWG (0.205 mm
) MINIMUM.
SIGNAL
TRANSMIT, RS232
RECEIVE, RS232
COM4 GROUND
2-49